Vertical Laminar Flow Cabinet LT-VLC212 Catalog

Overview

Labtro Vertical Laminar Flow Cabinet LT-VLC212 works by passing room air through a high-efficiency HEPA filtration system before directing purified air vertically across the SUS304 stainless steel work surface. The filtered airflow creates an ISO Class 5 / Class 100 clean working zone for contamination-sensitive sample handling. A centrifugal fan maintains stable airflow, while the intelligent five-stage control system allows air velocity adjustment from 0.2 to 0.6 m/s. The LCD controller manages airflow, sterilization timing, socket operation, front-window status, interlock functions, cumulative working time, and remote operation. A UVC sterilization lamp supports cabinet decontamination between procedures, with a lifetime reminder for maintenance. The tempered glass working door provides clear visibility, while adjustable castors allow convenient positioning within the laboratory.

Specifications:

Applicable Station One person, one side
Airflow Direction Vertical
Cleanliness Class ISO Class 5 / Class 100
Collection Efficiency ≥99.995% for 0.3 µm particles
Colony Count ≤0.5/vessel, Ø90 mm Petri dish
Wind Speed Range 0.2 to 0.6 m/s
Wind Speed Control Intelligent multi-level / five-stage adjustable
Noise Level ≤62 dB
Vibration Semi-Peak Value ≤3 µm
Illumination ≥300 lux at operation center
Maximum Work Surface Load 50 kg
Work Surface Material SUS304 stainless steel
Outer Shell Cold-rolled steel with electrostatic spray coating
Operation Door Tempered glass
HEPA Filter 820 × 600 × 50 mm, 1 pc
Fluorescent Lamp 15 W × 2
UVC Lamp 15 W × 2
Display LCD color screen
Sterilization Reservation Provided
Sterilization Timer Provided
Socket Control Controllable through LCD controller
Front Window Status Display Provided
Interlock Function Three-interlock setting
Cumulative Work Timer Provided
Remote Control Provided
Built-in Universal Sockets 2 pcs
UV Lamp Lifetime Alarm Provided
Adjustable Castors Provided
One-Key UV Control Provided
Internal Dimensions (W×D×H) 870 × 690 × 520 mm
External Dimensions (W×D×H) 1010 × 750 × 1600 mm
Packing Dimensions (W×D×H) 1160 × 872 × 1765 mm
Rated Power 250 W
Maximum Socket Load 300 W per socket
Power Supply AC 220 V, 50/60 Hz
Rated Current 1.1 A
Net Weight 130 kg
Gross Weight 170 kg
Features:
  • ISO Class 5 clean-air environment
    ↳ Maintains controlled particulate conditions during sensitive procedures

    Vertical HEPA-filtered airflow
    ↳ Directs purified air downward across the working area

    Five-stage airflow adjustment
    ↳ Provides selectable air velocity from 0.2 to 0.6 m/s

    LCD color-screen controller
    ↳ Centralizes airflow, sterilization, socket, timer and window-status functions

    UVC sterilization system
    ↳ Supports work-zone treatment between operating procedures

    SUS304 work surface
    ↳ Provides a corrosion-resistant and easy-to-clean workspace
Applications :
  • Microbiology Laboratories → Clean preparation of non-hazardous microbiological samples

    Cell Culture Laboratories → Supports clean handling of suitable non-hazardous cultures

    Pharmaceutical Laboratories → Contamination-sensitive sample and reagent preparation

    Research Laboratories → Clean experimental setup and material handling

    Quality Control Laboratories → Controlled preparation of test materials

    Universities → Clean-air workstation for laboratory teaching and research
Labtro